7400 Series 20KV/10KV AC High Withstand Voltage Tester
Model      
7470 7473 ALL
SPECIFICATIONS
MODEL 7470 7473
Output Rating 10KVAC/20mA 20KVAC/10mA
AC WITHSTAND VOLTAGE
  Range Resolution Accuracy
Output Voltage, ACV (7473 ONLY) 0 - 20000 10 ±(2% of setting + 20V)
Output Voltage, ACV (7470 ONLY) 0 - 10000 10 ±(2% of setting + 10V)
Output Frequency 50Hz/60Hz ±100ppm, User Selection
Output Waveform Sine Wave ,THD.<2% (Resistive Load), Crest Factor = 1.3 - 1.5
Output Regulation ± (1% of output + 10V), From no load to full load
SETTINGS
Hi / Lo -Limit AC Current, mA (7473 ONLY) 0 - 9.999 0.001 ±(2% of setting + 2counts)
Hi / Lo -Limit AC Current, mA (7470 ONLY) 0 - 9.999/10.00 - 20.00 0.001/0.01 ±(2% of setting + 2counts)
Ramp Up Time, second 0.1 - 999.9 0.1 ±(0.1% + 0.05sec)
Ramp Down Time, second 0.0 - 999.9 0.1
Dwell Time 0, 0.3 - 999.9 sec (0=continuous) 0.1
0, 0.1 - 999.9 min (0=continuous) 0.1
0, 0.1 - 999.9 hr (0=continuous) 0.1
Cycle Test 0 - 9999 Times, (0=continuous)
Arc Detection 1 - 9 ranges ( 9 is the most sensitiviy)
AC Current Offset, mA 0 - 2.000 or Auto Set ( Option )
MEASUREMENT
AC Voltage, KV (7473 ONLY) 0 - 20.00 0.01 ±(2% of reading + 2counts)
AC Voltage, KV (7470 ONLY) 0 - 10.00 0.01 ±(2% of reading + 2counts)
AC Current Range, mA 0-3.500 0.001 ±(2% of reading + 3counts)
3.00-10.00 (7473 ONLY) 0.01
3.00-20.00 (7470 ONLY) 0.01
GENERAL
Input Voltage AC 115/230VAC ± 15% , 50/60Hz ± 5% Fuse: 6.3A Slow
PLC Remote Control Input : Test , Reset , Memory1.2.3 ,Interlock
Output : Pass , Fail , Processing
Safety Smart-GFI Function (<1mA)
Memory 50 memories
Display 20*2 LCD with back light
Key Lock To prevent unauthorized alteration of the test Parameters
Calibration Build-in software and external calibrated meters
Interface RS232/GPIB /Printer Card (Option)
Alarm Volume Setting 10 ranges set by the numeric keys on the front panel
Environment 0-40℃, 20-80%RH
Dimension /Net Weight 430mm(W)×133mm(H)×400mm(D)/ 22.5Kg